Infant Incubator
Infant Incubator
- Product name: InfantCare 6 Plus
- Twins Holding Capability
- Microprocessor based servo controlled temperature system
- Control modes: Air mode and baby skin mode
- Humidity control and oxygen concentration control system
- Set temperature, air temperature, baby skin temperature, humidity,oxygen concentration, timer and heating power are displayed separately by LED
- Self - testing function, various failure alarms by audible and visual
- Alarm function: Power failure,Over temperature,Temperature deviation,Fan failure,Temperature sensor failure,Low water level, O2 concentration deviation
- 37° temperature set function
- Embedded integrated sensor module
- Triple protection for over temperature with separate cut off device, more safety system
- The inclination of infant bed is adjustable
- Double wall hood, 4 operating windows and 2 iris ports,5 infusion seals
- Independent locking device for front door
- RS232 connector, Oxygen inlet,1 tray
- Power supply AC110/220V 60/50Hz
- Power input 650V
- Operating condition:
- Environment temperature 20°C 30°C
Environment relative humidity 30 75
Environment air velocity of flow <0.3m/s
- Air temperature control range 25°C 37°C 37.1°C 38°C
- Skin temperature control range 32°C 37°C 37.1°C 38°C
- Humidity display range 0 100
- Humidity control range 40 95
- Oxygen concentration display range 0 100
- Oxygen concentration control range 20 60
- Oxygen concentration deviation alarm ±5
- Skin temperature sensor accuracy ±0.3°C
- Temperature fluctuation ±0.5°C
- Uniformity of mattress temperature ≤ 0.8°C
- Warm-up time(from 25°C) 30min
- Internal noise level 50dB(A)
- Infant bed tilt angle ±10°
- Mattress size 65cm(L)*37cm(W)
- Trough capacity 1000mL
- Air filter 0.5µm
Specification:
Infant Incubator
Infant Incubator
- Product name: InfantCare 6
- Twins Holding Capability
- Microprocessor based servo controlled temperature system
- Control modes: Air mode and baby skin mode
- Humidity control and oxygen concentration control system
- Set temperature, air temperature, baby skin temperature, humidity,oxygen concentration, timer and heating power are displayed separately by LED
- Self - testing function, various failure alarms by audible and visual
- Alarm function: Power failure,Over temperature,Temperature deviation,Fan failure,Temperature sensor failure,Low water level, O2 concentration deviation
- 37° temperature set function
- Embedded integrated sensor module
- Triple protection for over temperature with separate cut off device, more safety system
- The inclination of infant bed is adjustable
- Double wall hood, 4 operating windows and 2 iris ports,5 infusion seals
- Independent locking device for front door
- RS232 connector, Oxygen inlet,1 tray
- Power supply AC110/220V 60/50Hz
- Power input 650V
- Operating condition:
- Environment temperature 20°C 30°C
Environment relative humidity 30 75
Environment air velocity of flow <0.3m/s
- Air temperature control range 25°C 37°C 37.1°C 38°C
- Skin temperature control range 32°C 37°C 37.1°C 38°C
- Humidity display range 0 100
- Humidity control range 40 95
- Oxygen concentration display range 0 100
- Skin temperature sensor accuracy ±0.3°C
- Temperature fluctuation ±0.5°C
- Uniformity of mattress temperature ≤ 0.8°C
- Warm-up time(from 25°C) 30min
- Internal noise level 50dB(A)
- Infant bed tilt angle ±10°
- Mattress size 65cm(L)*37cm(W)
- Trough capacity 1000mL
- Air filter 0.5µm
Specification:
Infant Incubator
Infant Incubator
- Product name: InfantCare 4
- Microprocessor based servo controlled temperature system
- Control modes: Air mode and baby skin mode
- Humidity is adjustable continuously
- Set temperature, air temperature, baby skin temperature, timer and heating power
- are displayed separately by LED
- Integral timer to record therapy time
- Self - testing function, various failure alarms by audible and visual
- Alarm function: Power failure,Over temperature,Temperature deviation,Fan failure,Temperature sensor failure
- >37°C temperature set function
- Triple protection for over temperature with separate cut off device, more safety system
- The inclination of infant bed is adjustable
- Single wall hood, 4 operating windows and 2 iris ports,5 infusion seals
- Independent locking device for front door
- RS232 connector, Oxygen inlet, 1 tray
- Power supply AC110/220V 60/50Hz
- Power input 650V
- Operating condition:
- Environment temperature 20°C 30°C
Environment relative humidity 30 75
Environment air velocity of flow <0.3m/s
- Air temperature control range 25°C 37°C 37.1°C 38°C
- Skin temperature sensor accuracy ±0.3°C
- Temperature fluctuation ±0.5°C
- Uniformity of mattress temperature ≤ 0.8°C
- Warm-up time(from 25°C) 30min
- Internal noise level 50dB(A)
- Infant bed tilt angle ±10°
- Mattress size 65cm(L)*37cm(W)
- Trough capacity 1200mL
- Air filter 0.5µm
Specification:
Infant Incubator
Infant Incubator
- Product name: InfantCare 2
- Microprocessor based servo controlled temperature system
- Control modes: Air mode
- Humidity is adjustable in two levels
- Set temperature, air temperature, heating power are displayed separately by LED
- Self - testing function, various failure alarms by audible and visual
- Alarm function:
Power failure,Over temperature,Temperature deviation,Fan failure,Temperature sensor failure - >37°C temperature set function
- Triple protection for over temperature with separate cut off device, more safety system
- The inclination of infant bed is adjustable
- Single wall hood, 4 operating windows and 2 iris ports,3 infusion seals
- locking device for front door
- RS232 connector, 1 tray
- Power supply AC110/220V 60/50Hz
- Power input 650V
- Operating condition:
- Environment temperature 20°C 30°C
Environment relative humidity 30 75
Environment air velocity of flow <0.3m/s
- Air temperature control range 25°C 37°C 37.1°C 38°C
- Skin temperature control range 32°C 37°C 37.1°C 38°C
- Humidity display range 0 100
- Humidity control range 40 95
- Oxygen concentration display range 0 100
- Skin temperature sensor accuracy ±0.3°C
- Temperature fluctuation ±0.5°C
- Uniformity of mattress temperature ≤ 0.8°C
- Warm-up time(from 25°C) 30min
- Internal noise level 50dB(A)
- Infant bed tilt angle ±10°
- Mattress size 65cm(L)*37cm(W)
- Trough capacity 1200mL
- Air filter 0.5µm
Specification:
Transport Incubator
Transport Incubator
- Product name: AmpCare 1
- Air mode and baby mode controlled by micro-computer
- Use AC Power and DC Power alternatively,DC 12V or DC 24V Power on the ambulance
- Set temperature,air temperature,skin temperature and internal battery power are displayed by LED
- Self-testing function,various failure alarms by audible and visual
- Alarm function:
Power failure,Over temperature,Temperature deviation,Fan failure,Temperature sensor failure - Double wall hood with side door and front door,4 operating windows and 2 iris ports,5 infusion seals
- The infant bed can be pulled out from side door easily
- Natural air flow humidity
- The height od whole unit is adjustable
- Oxygen supply system
- LED observe light,adjustable brightness
- Build-in storage cabinet
- Power supply AC110/220V 60/50Hz
- Power input 450V
- Operating condition:
- Environment temperature 10°C 30°C
Environment relative humidity 30 75
Environment air velocity of flow 1m/s
- Air temperature control range 25°C 37°C 37.1°C 38°C
- Skin temperature sensor accuracy ±0.3°C
- Skin temperature control range 25°C 37°C 37.1°C 38°C
- Air/Skin temperature sensor accuracy ±0.3°C
- Temperature fluctuation ±1°C
- Uniformity of mattress temperature ≤ 1.5°C
- Air temperature deviation alarm ±3°C
- Skin temperature deviation alarm ±1°C
- Warm-up time(from 25°C) 40min
- Internal noise level 60dB(A)
- Main battery DC12V 40AH,Li-ion battery
- Charge time 12hours
- Battery use time 2hours
- Auxiliary battery DC8.4V 0.17AH,Li-ion battery
- Mattress size 63cm(L)*32cm(W)
- Trough capacity 1200mL
- Air filter 0.5µm
Specification:
Infant Incubator & Radiant Warmer
Infant Incubator
- Product name: MultiCare-OP
- Incubator and Warmer mode, servo controlled by micro-computer, convert automatically
- Multi audible and visual alarm finctions, self-testing finction
- 10.2" color LCD display touchscreen with 24h parameter graph trends
- Electrical lifting canopy to convert to warmer mode from incubator mode
- LED observe light with brightness adjustable
- Infant bed can be pulled out bidirectionally
- Foor pedal electrical height adjustment system for optimum adaptation to clincal users
- Electrical infant bed tiliting system
- Built-in baby scale system, with real-time graph
- Downside phototherapy unti
- SPO2 monitor system
- 37° temperature set function
- Triple protection for ober temperature with separate cut off device, more safety system
- Double fan, air curtain strucure, better temperature uniformity , reduce heat loss during operation
- Double wall hood, 4 operating windows, 1 iris port, 9 big infusion seals, independent locking device
- Oxygen inlet
- One tray and one big drawer
- Data stroage interface and RS232 data interface
- Power supply AC110/220V 60/50Hz
- Power input 1000VA
- Operating condition:
- Environment temperature 20°C 30°C
Environment relative humidity 30 75
Environment air velocity of flow 0.3m/s
- Air temperature control range 25°C 37°C 37.1°C 39°C
- Skin temperature control range 32°C 37°C 37.1°C 39°C
- Skin temperature sensor accuracy ±0.2°C
- Temperature fluctuation ±0.5°C
- Uniformity of mattress temperature ≤ 0.8°C
- Humidity display range 0% - 99%
- Humidity control range 30% - 95%
- Oxygen Concentration display range 0% - 100%
- Oxygen Concentration control range 20% - 60%
- Warm-up time (from 25°C) <30min Infant Radiant Warmer Mdoe (Open Mode)
- Skin temperature control range 32°C 38°C
- Uniformity of mattress temperature ≤ 2°C
- Warm-up time(from 25°C) 45min
Specification:
Infant Radiant Warmer
Infant Radiant Warmer
- Product name: OpenCare 2
- Microprocessor based servo controlled temperature system
- Three modes: pre-warm, manual or automatic
- Audible and visual Alarm functions : Power failure, temperature deviation, temperature sensor failure, skin temperature failure, over temperature
- 8 inch LCD colorful touch screen
- Carbon steel heater more effective
- Special design inclined heater head is more convenience for X ray test image
- Digital skin temperature sensor,the data is accurate and stable
- Alloyed bracket,fastness
- Integral timer to record therapy time
- Bigger mattress
- Skin temperature trend for 2 hours
- Electric infant bed tilt angle is ±12°
- Four transparent protector folded with 2 infusion seals
- Convenient X-ray cassette tray under infant bed
- Electric height adjusted suit for operation
- With observe light,oxygen supply system and LED phototherapy
- With RS-232 connector,2 trays and 2 drawers
- Power supply AC110/220V 60/50Hz
- Power input 1100V
- Operating condition:
- Environment temperature 18°C 30°C
Environment relative humidity 30 75
Atmospheric Pressure 700~1060hpa
Environment air velocity of flow < 0.3m/s
- Skin temperature control range 32°C ~ 38°C
- Temperature control accuracy ≤0.5°C
- Skin temperature sensor accuracy ±0.3°C
- Mattress temperature uniformity ≤2°C
- Infant Bed tilt angle ±12°
- Warm-up time(from 25°C) ≤ 30min
- LED observation light 3W
- Distance from heater to mattress 80cm
- Mattress size 78cm(L)*64cm(W)
- Timer range 0~99h59min Height adjustable range 0~20cm
- LED phototherapy light 36cm 620-2800µw/c ㎡
- LED phototherapy light 50cm 400-1620µw/c ㎡
- Oxygen cylinder 6L X 2
- Oxygen flow meter 0-10L/min
Specification:
Infant Radiant Warmer
Infant Radiant Warmer
- Product name: OpenCare 1 PLUS
- Microprocessor based servo controlled temperature system
- Three modes: pre-warm, manual or automatic
- Set temperature and skin temperature display separately
- Special design inclined heater head is more convenience for X rays
- Carbon steel heater for more effectivity
- Audible and visual Alarm functions : Power failure, temperature deviation, temperature sensor failure, skin temperature failure, over temperature
- APGAR timer to record therapy time
- Convenient X-ray cassette under infant bed
- Electric inclination of infant bed
- LED observation light adjustable,brightness adjustable
- Four Transparent protector folded with 2 infusion seals
- LED phototherapy unit,Oxygen supply system
- Bigger infant bed
- Built in low pressure aspirator
- With RS-232 connector,2 trays and 2 drawers
- Power supply AC110/220V 60/50Hz
- Power input 1100V
- Operating condition:
- Environment temperature 18°C 30°C
Environment relative humidity 30 75
Atmospheric Pressure 700~1060hpa
Environment air velocity of flow 0.3m/s
- Skin temperature control range 32°C 38°C
- Skin temperature control accuracy ±0.5°C
- Skin temperature sensor accuracy ±0.3°C
- Mattress temperature uniformity 2°C
- Temperature fluctuation ±1°C
- Mattress temperature uniformity 2°C
- Infant Bed tilt angle ±12°
- Warm-up time(from 25°C) 30min
- Heater head horizontal ±90°
- LED observation light 3W
- Distance from heater to mattress 80cm
- Mattress size 78cm(L)*64cm(W)
- Timer range 0~99h59min
- LED phototherapy light 36cm 620-2800µw/c ㎡
- LED phototherapy light 50cm 400-1620µw/c ㎡
- Low pressure aspirator:
- Max vacuum adjustable ≥-25kpa(-188mmHg)
Storage bottle 1000mL
Oxygen cylinder 6L X 2
Oxygen flow meter 0-10L/min
Specification:
Infant Radiant Warmer
Infant Radiant Warmer
- Product name: OpenCare 2
- Microprocessor based servo controlled temperature system
- Three modes: pre-warm, manual or automatic
- Audible and visual Alarm functions : Power failure, temperature deviation, temperature sensor failure, skin temperature failure, over temperature
- 8 inch LCD colorful touch screen
- Carbon steel heater more effective
- Special design inclined heater head is more convenience for X ray test image
- Digital skin temperature sensor,the data is accurate and stable
- Alloyed bracket,fastness
- Integral timer to record therapy time
- Bigger mattress
- Skin temperature trend for 2 hours
- Electric infant bed tilt angle is ±12°
- Four transparent protector folded with 2 infusion seals
- Convenient X-ray cassette tray under infant bed
- Electric height adjusted suit for operation
- With observe light,oxygen supply system and LED phototherapy
- With RS-232 connector,2 trays and 2 drawers
- Power supply AC110/220V 60/50Hz
- Power input 1100V
- Operating condition:
- Environment temperature 18°C 30°C
Environment relative humidity 30 75
Atmospheric Pressure 700~1060hpa
Environment air velocity of flow < 0.3m/s
- Skin temperature control range 32°C ~ 38°C
- Temperature control accuracy ≤0.5°C
- Skin temperature sensor accuracy ±0.3°C
- Mattress temperature uniformity ≤2°C
- Infant Bed tilt angle ±12°
- Warm-up time(from 25°C) ≤ 30min
- LED observation light 3W
- Distance from heater to mattress 80cm
- Mattress size 78cm(L)*64cm(W)
- Timer range 0~99h59min Height adjustable range 0~20cm
- LED phototherapy light 36cm 620-2800µw/c ㎡
- LED phototherapy light 50cm 400-1620µw/c ㎡
- Oxygen cylinder 6L X 2
- Oxygen flow meter 0-10L/min
Specification:
Infant Phototherapy Unit
Infant Phototherapy Unit
- Product name: PH-200
- LED Blue bulb as the radiant source, irradiating uniformly and efficient.
- Circle 360 double sides for more effective radiantion
- Small in size, light and high irradiation
- Two phototherapy modes:up side phototherapy and down side phototherapy,can be use separately
- Therapy time of phototherapy units up and down side are separately display
- Automatically record each therapy time and total used therapy time
- Air temperature and skin temperature separately display
- Power off memory, alarm silence key , locking key and self-examination function
- Four alarm functions: Power failure, probe failure, over temperature, fan failure
- Count down setting avilable
- Two sides observation parts
- Air spring structure make the hood is easy open and close.
- The aluminum alloy is applied on the base ,with 2 drawers and 1 tray
- Power supply AC220V±10% 50Hz±2%
- Power input ≤400VA
- Blue light wavelength: 420nm~490nm
- Circle 360 light radiation: 3700µW/c㎡
- Up side light radiation: 3300µW/c㎡
- Down side light radiation: 2400µW/c㎡
- Life span of blue light tube: ≥20000hours
- Air temperature display range: 0~45.0°C
- Skin temperature display range: 0~45.0°C
- Time accuracy: 1min/2h
- Count-up time range: 0~99.9 hours
- Count-down time range: 1min~99.9hours
- Total cumulative time: 0~999999hours
- Mattress size: 613X300mm
- Working environment:
- Ambient temperature t: +18°C +30°C
Relative humidity: 30% 75%
Atmospheric pressure: 700hpa~1060hpa
Specification:
Infant Phototherapy Unit
Infant Phototherapy Unit
- Product name: PH-10
- Equal light distribution
- High intensity
- Quiet, no noise
- Five blue fluorescent lights
- The angle of head and height are adjustable
- Steel support pole, four castors with brake
- Integral time record therapy time
- Power supply AC110/220V 60/50Hz
- Power input 120VA
- Blue light wavelength: 400nm~550nm
- Total Irradiance for bilirubin 1300µW/c#13217;
- Height Adjust Range 1100~1600mm
- Pitching angle of phototherapy head ±90°
- Time Accurancy 1min/12h
- Integral time range 0h 9999h 59min
- Working environment:
- Ambient temperature t: +18°C +30°C
Relative humidity: 10% 85%
Atmospheric pressure: 700hpa~1060hpa
Specification:
Follow Us
Email Updates
Join our digital mailing list and get news
special deals and latest catalogs